GeneRotex 96 Nucleic Acid Extractor

The Tianlong GeneRotex 96 nucleic acid extraction system is an automated nucleic acid extraction system featuring our innovative rotary mixing technology (RMT), which can significantly reduce aerosol generated during the purification process, minimize the risk of false positive results caused by cross-contamination, and increase the accuracy of experiment results. The GeneRotex 96 Tianlong nucleic acid extractor is compatible with 96 deep-well plates and six tube strips, making this nucleic acid extraction system capable of offering high-throughput extraction and reducing waste without the conventional inconveniences caused by fixed throughput. Flexible and efficient, the GeneRotex 96 nucleic acid extractor offer 1 to 96 samples per run.

Features and specifications.

A1.

7 inch color touch screen: Built-in 7-inch full-color LCD screen, easy to operate the experiment.
High throughput and efficient extraction: The Tianlong GeneRotex 96 nucleic acid extraction system is a 6*16 extraction module with a special 96 deep well plate and 6 strip tube designed to ensure high throughput and reduce reagent waste. The GeneRotex 96 allows you to extract from 1 to 96 samples per run.

B1.

Innovative rotary mixing technology: Based on Tianlong’s innovative rotary mixing technology(RMT), GeneRotex 96 nucleic acid extractor can reduce aerosol generated during the experiment and minimize the risk of false positives caused by cross-contamination while being super quiet during operation. 

Negative pressure system with HEPA filtration: Featuring a negative pressure ventilation design, the GeneRotex96 automated nucleic acid extraction system is available for replaceable HEPA filtration which ensure the exhausted air has no biological hazards.

High purification and reliable results: Less than 1% residual amount of magnetic beads increase confidence in your experiment result.

Model

GeneRotex 96 

Throughput

1-96

Reaction Volume

30-1000uL

Sample Volume

200uL

Compatible Consumables

96-deep-well plates
customized 6 strip tube

Inter-well Extraction Difference

CV ≤3%

Rotational Speed

≤3000rpm

Heating Temperature

Lysis heating: room temperature to 120°C
Elution heating: room temperature to 120°C

Mixing Method

Rotary mixing

Operation Mode

7-inch full-color LCD touch screen operation

Program Storage

Up to 1000 programs can be storaged

Protocol Management

Flexible to create, edit and delete protocols

Automatic Control

Motor-driven automatic opening and closing of the experiment cabinet

Magnetic Bead Residue

≤1%

Power Failure Protection

Choose freely whether or not to continue the experiment when the power is on again after cutting off

Contamination Control

Negative pressure HEPA exhaust filter module;
Built-in UV disinfection module

Connection Port Type

USB port

Weight

45kg (net)

Instrument Dimensions

510mm*490mm*480mm (W*L*H)

Power Supply and Power Consumption

AC 100V-240V, 50/60HZ; 600VA

Certificate

IVDR CE